These data were collected with a Sea-Bird CTD (conductivity, temperature, depth) instrument - a SBE19plusV2 SeaCAT Profiler CTD. This instrument was deployed on the USGS SEABed Observation and Sampling System (SEABOSS). The SEABOSS is an instrument developed by the USGS to collect samples and video and photographic images of the seabed to aid in the interpretation of seabed sonar imagery. The CTD is mounted horizontally on the SEABOSS frame and its maximum depth is approximately 1 meter above the seabed. The specifications for the CTD instrument are as follows:
SBE19plusV2, manufactured by Sea-Bird Electronics, Inc.
Firmware version: 2.2c
Firmware date: 7 December 2009
Manufactured date: 29 July 2010
Software version: 2.3.0
Internal sensors:
main temperature, serial number 01906617
main conductivity, serial number 01906617
main pressure, type, strain, serial number 3161799
External sensors from WET Labs:
ECO-FLNTURT CHL, serial number FLNTURT-1798; WET Labs fluorometer, ECO-AFL/FL
ECO-FLNTURT TURB, serial number FLNTURT-1798; WET Labs turbidity meter, ECO-NTU
Configuration:
Instrument battery operated; not connected to a conducting cable
Profile mode: sampling rate 4 per second, averaged to provide each data point
CTD deployed on the USGS SEABed Observation and Sampling System (SEABOSS)
Data archive:
All data; downcast, drift, and upcast
Sensor specifications:
Temperature (degrees C):
measurement range -5 to +35 degrees C; initial accuracy +-0.005; resolution 0.0001;
Conductivity (S/m):
measurement range 0 to 9 S/m; initial accuracy +- 0.0005; resolution varies: 0.00005 (most oceanic water; resolves 0.4 ppm in salinity); 0.00007 (high salinity water; resolves 0.4 ppm in salinity); 0.00001 (fresh water; resolves 0.1 ppm in salinity)
Strain gauge pressure:
measurement range 0 to full scale range: 20/100/350/600/1000/2000/3500/7000; initial accuracy +- 0.1% of full scale range; resolution 0.002% of full scale range.
Turbidity sensor:
wavelength 700 nm; sensitivity 0.01 NTU, rant 0-25 NTU
Fluorescence sensor:
Chlorophyll-a 470 nm EX/695 nm EM; sensitivity 0.2 micro-g/l; range 0-125 micro-g/l Chl